Synergy H4 Micro Plate Reader
Synergy H4 Micro Plate Reader location: AS8-2028 status: in service reserve use
Multimode plate reader with dual injectors for time-critical reagent addition. Handles plates up to 1536 wells. Modes include luminescence, UV-visible absorbance, fluorescence intensity, time resolved fluorescence, fluorescence polarization, fluorescence resonance energy transfer (FRET), bioluminescence resonance energy transfer (BRET), and alphascreen. Dual monochrometers for excitation and emission provide high spectral resolution. Filter mode for high optical throughput
when required.
Specifications
General | ||
Wavelength Selection | Hybrid Technology™ (patent-pending) Quadruple Monochromators and Filters / Dichroics | |
Detection method | Fluorescence, Time-Resolved Fluorescence, Fluorescence Polarization,AlphaScreen/AlphaLISA®, Luminescence, UV-Visible Absorbance |
|
Read method | End point, kinetic, spectral scanning, well-area scanning | |
Microplate types | Monochromator system 1- to 384-well plates Filter system 1- to 1536-well plates PCR plates Compatible with Take3™ Multi-Volume Plate with 2 µL microspots |
|
Temperature control | 4° C above ambient to 65°, C + 0.5° C at 37° C | |
Shaking | Yes | |
Software | Gen5™ included; Control through USB or serial port | |
Automation | Yes | |
AlphaScreen/AlphaLISA | ||
Light source | Tungsten Halogen Lamp | |
Sensitivity | 100 amol of biotinylated-LCK-P peptide, 25 µL/well in 384-well plate | |
Dynamic range | >6 decades | |
Reading speed | 2 minutes typical for 96-well plate | |
Detection system | Ultra Low Noise PMT | |
Absorbance | ||
Light source | Xenon Flash Lamp | |
Wavelength selection | Monochromator | |
Wavelength range | 230 - 999 nm, 1 nm increment | |
Bandpass | 2 nm (230-285 nm), 4 nm (>285 nm) | |
Dynamic range | 0 - 4.0 OD | |
Resolution | 0.0001 OD | |
Pathlength correction | Yes | |
Monochromator wavelength accuracy | +2 nm | |
Monochromator wavelength repeatability | +0.2 nm | |
OD accuracy | <1% at 2.0 OD typical <3% at 3.0 OD typical | |
OD linearity | <1% from 0 to 3.0 OD typical | |
OD repeatability | <0.5% at 2.0 OD typical | |
Stray light | 0.03% at 230 nm typical | |
Reading speed | 96: 11 seconds 384: 22 seconds |
|
Fluorescence Intensity | ||
Sensitivity | Monochromators: Fluorescein 2 pM typical (0.2 fmol/well; 384-well plate) (Top) Fluorescein 2.5 pM typical (0.25 fmol/well; 384-well plate) (Bottom) Filters/mirrors: Fluorescein 1 pM typical (0.1 fmol/well; 384-well plate) |
|
Light source | Tungsten Halogen Lamp High Energy Xenon Flash Lamp |
|
Wavelength selection | Double grating monochromators (Top/Bottom) and, Deep blocking filters / dichroic mirrors (Top) |
|
Wavelength range | Monochromators: 250 - 900 nm Filters: 200 - 700 nm (900 nm option) |
|
Dynamic range | Monochromators: 5 decades Filters/mirrors: >6 decades |
|
Bandpass | Monochromators: variable 9, 13.5, 17, 20 nm Filters: filter-dependant, from 5 nm to >100 nm |
|
Detection system | Red shifted PMT for monochromator system Low Noise PMT for filter-based system |
|
Reading speed | 96: 11 seconds 384: 22 seconds 1536: 42 seconds |
|
Luminescence | ||
Sensitivity | 10 amol ATP typical (flash) | |
Wavelength range | 300 - 700 nm | |
Dynamic range | >6 decades | |
Fluorescence Polarization | ||
Light source | Tungsten Halogen | |
Sensitivity | 3 mP at 1 nM fluorescein typical | |
Wavelength selection | Deep blocking filters / dichroic mirrors (Top) | |
Wavelength range | 400 - 700 nm (300 - 900 nm option) | |
Time-Resolved Fluorescence | ||
Light source | High Energy Xenon Flash Lamp | |
Sensitivity | Europium 60 fM typical with filters (6 amol/well in 384-well plate) | |
Wavelength selection | Deep blocking filters / dichroic mirrors (Top) Double grating monochromator (Top / Bottom) |
|
Wavelength range | Filters: 200 - 700 nm (900 nm option) Monochromators: 250 - 900 nm |
|
Reagent Dispensers | ||
Number | 2 syringe pumps | |
Dispense volume | 5 - 1000 µL in 1 µL increment | |
Dead volume | 1.1 mL, 100 µL with back flush |