Here are some typical setup for Low-Dose condition.
Search: Mag=2650X(3000X with screen up), Spotsize=3-5, image mode
Alternatively, at 2650-3400x, switch to Diff. mode by pressing "Diff" button. Adjust camera length to 680mm using "Mag" knob. Focus the diffraction spot using
knob, and then adjust "defocus" until the image inside central spot expands to proper size. Personally, I like to "defocus" to clockwise side. What you got is a shadowing image inside of defocused diffraction central beam. The advantage of this is high contrast compared to a normal image mode.Focus: Mag=175kX, Spotsize=6. When work with tilting stage, make sure S1 or S2 sits on tilting axis of the geonimeter, the angles will posted on the panel.
Record: Mag=53000X(60kX), spotsize=3-5