9. Insert specimen holder into column

9. Insert specimen holder into column

Warning:

Be careful! Only at this stage, you might damage the scope or specimen holder mechanically. Be sure that you understand what you are doing. Should you feel any confusion about this procedure, please stop and ask for help.

Note:

For the sake of filament crystal, it is REQUIRED to turn down the filament to 10 before inserting the specimen rod. That way, in case IGP shoots high, there will be no subtancial damage to the LaB6 tip crystal. In general, filament should be kept at 10 or completely off until IGP recovers to below 26.

The procedure to insert a room temperature specimen rod is below.

This is an official Page of the University of Massachusetts Worcester Campus
Electron Microscopy Facility • 55 Lake Avenue North Worcester, Massachusetts 01655-0160
Questions or Comments? Email: gregory.hendricks@umassmed.edu Phone: 508-856-2602